| Power | Mains 100-245 V AC/50-60Hz, 400W, CE conform |
| Dimensions | 22 x 40 x 38 cm (W x D x H) |
| Display | LCD 4 x 40 digits with backlight illumination |
| Fermentor vessel | Pyrex glass with 5 to 8 side necks 0.3,0.4,1,3,5 l |
| Temperature control | High efficiency 150 W radiation heat source with gilded reflector |
| Regulation | from 5°C over RT to 70°C |
| Measurement | from 0 to 99.9°C in 0.1°C steps |
| Precision | +/- 0.2°C (0 to 60°C) |
| Sensor | Pt 100 incorporated in the glass electrode of the pH probe |
| pH control | sterilisable pH electrode pH 0-13 with nano-suspension reference electrolyte (high signal stability), automatic temperature correction, two-point semiautomatic calibration and Variopin connector |
| Resolution | 0.01 pH unit |
| Precision | +/- 0.02 pH unit |
| pO2 control | sterilisable Clark type oxygen sensor with fast response, automatic temperature correction, two-point semiautomatic calibration, dissolved oxygen (DO) control through regulation of the airflow rate |
| Range | 0 to 25 mg oxygen/ l, in 0.1 mg/l steps |
| Air flow | 0 to 5 l/min. in 0.1 l/min. steps, measured by mass flow meter, linearity +/- 3%,reproducibility +/- 0.5% |
| Control | proportional valve controlled by microprocessor |
| Supplied air pressure | 0,05 – 0,2 Mpa (0,5 - 2 atm) |
| Agitation | 40 W Vibromixer 0 to 20 Hz (0 to 1200 rpm) with 1 or more stirring discs,Sterility similar to magnetic coupling |
| Selectable parameter | an additional parameter can be controlled by the instrument (foaming control, pCO2, optical density, conductivity, etc.); with standard 0-10V or 0-20mA output |
| Ports | One large quadruple sampling or additions port with four needles with LAMBDA PEEK double-seal connections, used for sampling, inoculation, antifoam, feeds, harvest, addition of correction solutions etc., additional double ports are available. |
| Pumps | up to 4 independent pumps with speed variation from 0 to 100 % can be used with Minifor (PRECIFLOW, MULTIFLOW or HiFLOW) |
| Gas flow control | In addition to pumps, several electronic flow controllers with flow rate ranges of 0-5 l/min (MASSFLOW 5000) or 0-500 ml/min (MASSFLOW 500) can be used for the controlled addition of gases (e.g. N2, O2, air, CO2) in cell cultures. |
| Working temperature | 0 – 40 °C |
| Working humidity | 0 - 90 % RH, not condensing |
| Weight | 7.5 kg |
| PC control | complete PC control and data processing using the fermentation software FNet (for up to 6 MINIFOR fermenters) or SIAM (for an even higher number of instruments) |