Specifications

Bioreactor Technical Specification

ParametrValue
PowerMains 100-245 V AC/50-60Hz, 400W, CE conform
Dimensions22 x 40 x 38 cm (W x D x H)
DisplayLCD 4 x 40 digits with backlight illumination
Fermentor vesselPyrex glass with 5 to 8 side necks 0.3,0.4,1,3,5 l
Temperature controlHigh efficiency 150 W radiation heat source with gilded reflector
Regulationfrom 5°C over RT to 70°C
Measurementfrom 0 to 99.9°C in 0.1°C steps
Precision+/- 0.2°C (0 to 60°C)
SensorPt 100 incorporated in the glass electrode of the pH probe
pH controlsterilisable pH electrode pH 0-13 with nano-suspension reference electrolyte (high signal stability), automatic temperature correction, two-point semiautomatic calibration and Variopin connector
Resolution0.01 pH unit
Precision+/- 0.02 pH unit
pO2 controlsterilisable Clark type oxygen sensor with fast response, automatic temperature correction, two-point semiautomatic calibration, dissolved oxygen (DO) control through regulation of the airflow rate
Range0 to 25 mg oxygen/ l, in 0.1 mg/l steps
Air flow0 to 5 l/min. in 0.1 l/min. steps, measured by mass flow meter, linearity +/- 3%,reproducibility +/- 0.5%
Controlproportional valve controlled by microprocessor
Supplied air pressure0,05 – 0,2 Mpa (0,5 - 2 atm)
Agitation40 W Vibromixer 0 to 20 Hz (0 to 1200 rpm) with 1 or more stirring discs,Sterility similar to magnetic coupling
Selectable parameteran additional parameter can be controlled by the instrument (foaming control, pCO2, optical density, conductivity, etc.); with standard 0-10V or 0-20mA output
PortsOne large quadruple sampling or additions port with four needles with LAMBDA PEEK double-seal connections, used for sampling, inoculation, antifoam, feeds, harvest, addition of correction solutions etc., additional double ports are available.
Pumpsup to 4 independent pumps with speed variation from 0 to 100 % can be used with Minifor (PRECIFLOW, MULTIFLOW or HiFLOW)
Gas flow controlIn addition to pumps, several electronic flow controllers with flow rate ranges of 0-5 l/min (MASSFLOW 5000) or 0-500 ml/min (MASSFLOW 500) can be used for the controlled addition of gases (e.g. N2, O2, air, CO2) in cell cultures.
Working temperature0 – 40 °C
Working humidity0 - 90 % RH, not condensing
Weight7.5 kg
PC controlcomplete PC control and data processing using the fermentation software FNet (for up to 6 MINIFOR fermenters) or SIAM (for an even higher number of instruments)


(c) 2008 Lambda Laboratory Instruments | redakční systém